Decoding the error messages
Sample error message
Release CoaLab V1.1 (2.26) from Jun 11 1999
Controller COALAB HBA V1.00 / THAU RTX51 V3.20
System error -3: Timeline execution error
00 15 80 87 04 43 88 03 61 89 04 6C
line a) System error -3: Timeline execution error
line b) 00 15 80 87 04 43 88 03 61 89 04 6C
a) The first line describes the
error class. The different classes are listed below.
Fatal Error -2: Invalid
command from module
Fatal Error -3: Timeline
Execution Error
Fatal Error -4: Result is
already measured (Error causes are corrected in software rev. 2.0)
Fatal Error -5: Results not in
timeline (Error causes are corrected in software rev. 2.0)
Fatal Error -6: No answer from
module
Fatal Error -7: Internal Error
in module
Fatal Error -9: Module is reset
Fatal Error -10: Module not
found
b) The second
line contains more precise information about the error. In most cases it
provides the information on which error type occurred, the command which
was sent and the error code.
00 Communication Error
01 Parameter Error
02 Load Error occurs during
program loading)
03 System error
04 Execution Error (generated
while action will start or action in process)
05 Request for cover switches
status
Error codes and their meaning are summarized in tables on the following
links.
|