Instrument line for medical diagnostics

American LAbor
Home
Contact us
About American LAbor
How to place an order
Sales terms & conditions

Disposables & prices
Accustasis 1000, AS 1000
Accustasis 2000, AS 2000
CoaData 1000, CD 1000
CoaData 2000, CD 2000
CoaData 3000, CD 3000
CoaSystem
CoaScreener
CoaLab, Nycomatic, BFA
Disposable details
Price list & order form

Other items for sale
New coagulation analyzers
Refurbished or used
Research analyzers
Pipettors
Printers

Service & FAQ
Accustasis 1000, AS 1000
Accustasis 2000, AS 2000
CoaData 1000, CD 1000
CoaData 2000, CD 2000
CoaData 3000, CD 3000
CoaSystem
CoaScreener
CoaLab, Nycomatic, BFA

If you have one of the above instruments and are no longer using it please contact us, we do sometimes buy units to refurbish and resell.



How to trouble shoot.

Accustasis 2000 and CoaData 2000 analyzers

GENERAL:
Appliances such as refrigerators, centrifuges, and other motorized appliances should NOT be on the same circuit with CoaCalc, CoaData, Accustasis or other computerized equipment. We suggest that this analyzer be connected to a surge protected electrical circuit or a UPS device, to protect it’s sensitive electrical components.

If your results are inconsistent the optics may have become contaminated by fluids being pipetted directly into an empty measuring well, without a cuvette in place. Turn off the analyzer. The measuring well optics must be clean. DO NOT spray any liquids onto or into the analyzer. Never use acetone, alcohol or any petroleum based products to clean this piece of equipment.

Cleaning may be done by moistening a q-tip with water and a mild soap, such as a dishwashing liquid, and wiping it around in a circular motion in the bottom of the test well. DO NOT saturate the q-tip. After using the soap solution, clean the soap solution by using only water on a q-tip, then use just a dry q-tip to remove the moisture. Turn on your analyzer and test with QC reagents.

SENS LIGHT stays on:
If you turn on your analyzer and the SENS light located on the keypad module will not go out, neither of the two channels will reset to zero, it is most likely the power supply board in the analyzer has quit working. Sometimes this is all that dies. However on rare occasions other components do not survive. This unit should come in to be repaired contact us.

STIR MOTOR:
Check each measuring well with a clean cuvette & stir bar to see if the measuring well stir motor is working. This is done by having the instrument on and warmed up, press reset to clear the display. Slide the spring loaded cover back to expose the measuring well. Place the cuvette into the measuring well, the timer for that channel should begin to count up and the mixer in the bottom of the cuvette should be spinning. If this did not occur contact us.

WILL NOT ZERO:
If the one channel resets and the other channel does not there may be a bulb burned out. This may be checked by blocking the light path in the measuring channel with a dry q-tip. With the instrument powered on slide the measuring channel cover back and while looking down into the well insert the q-tip. You should see the end of it light up as the intensity of the lamp is increased. If you do not see it illuminate contact us, this is not usually a repair that can be done in the field.

TEMPERATURE:
If your unit will not maintain 37 degrees it can be adjusted in your lab. You will need a mercury or traceable digital thermometer, a #1 and #2 flat blade screwdriver. Simply t
urn off the analyzer remove all chemicals from the heating block. Gently turn the analyzer over with the bottom up. Remove the outer corner screws indicated by blue circles, DO NOT REMOVE OTHER SCREWS, see figure 1 below.


Figure 1

Restore the unit to the upright operating position. With caution connect the power cord and turn unit on. Take a cuvette insert the thermometer, add ample water to cover the end of the thermometer, place in incubation block as close to center as possible, see figure 2.


Figure 2

The temperature is adjusted by turning a screw on an electronic device known as a potentiometer also referred to as a pot. Using a small #1 flat blade screwdriver the temperature may be adjusted up by turning the pot clockwise, or if necessary turn the temperature down by adjusting the pot counter clockwise. See figure 3 below for correct adjustment screw location marked by a red arrow.


Figure 3             Click here for a closer view.

The temperature adjustment should be made in small increments, for example 1/4 turn of the pot at a time. Then allow the temperature to stabilize for 20 - 30 minutes before any other adjustments. After the temperature is stable, turn off the analyzer,  replace the cover and secure with screws.

Please see service options for service charges.