CoaCalc and printer. Problem: it will not print results. Turn off CoaCalc (by unplugging the adapter), printer and Analyzer. 1. Turn on CoaCalc 2. Switch the printer on. 3. Then turn on analyzer. The analyzer must be turned on last for results to transfer properly. Does the startup information print when power is first applied? To test
unplug the power module for the CoaCalc for about 10 seconds If it did not print, is the power led on
the printer illuminated? If the above test failed to produce results contact us. Cause If someone
unplugs the adapter to the printer and does not turn off the switch on the
right side of the printer this will also discharge the batteries in the
printer. Check all connections on the
rear of the CoaCalc. There should also be a cable connecting the CoaCalc to the printer for data. There should be a cable connecting the analyzer to the CoaCalc. Also if there is a possibility someone has tampered with the unit check the row of switches on the rear of the printer and make sure they are in the positions indicated in the picture, all up except # 5 should be down. If these switches are changed the printer (ONLY) will need to be powered off and back on. Problem: CoaCalc resets itself every few seconds. This will happen when the printer obtains it's power from the CoaCalc by means of a jumper from the CoaCalc power output to the printer power input and the batteries in the printer have become discharged. Solution replace the battery
pack in the printer and allow it to charge overnight before using it.